LSPEC-YZ-600 Integrated Flame-Graphite Furnace Atomic Absorption Spectrophotometer
Product advantages
1. Total internal reflection optical system, full-band achromatic correction, automatic wavelength scanning and peak detection.
2. Features deuterium lamp subtraction background and self-absorption subtraction background.
3. Uniquely designed all-titanium burner ensures stable combustion and rapid thermal equilibrium.
4. Automatic adjustment of negative high voltage and lamp current, automatic positioning of optimal element lamp location, automatic ignition, and automatic flame-out protection.
5. 8-lamp rack (with optional 2 high-performance lamps) automatic control system. Enables comprehensive adjustment of hollow cathode lamps to optimal positions. Allows individual setting of specific element lamp positions. Permits free configuration of working lamps and preheating lamps, maxim ising savings in element lamp preparation time.
6. Graphite furnace heating: Users may freely define heating profiles as required.
7. Advanced constant-power temperature control design for graphite furnace: Ensures uniform graphite tube temperature distribution and maintains excellent dynamic stability at operating temperatures, guaranteeing analytical accuracy.
Product Parameters
Optical System
| Wavelength range | 185nm—910nm Auto-adjustment |
| Raster line density | 1800 lines/mm Aberration-corrected Czerny-Turner monochromator |
| Monochromator | Multi-range automatic switching: 0.1, 0.2, 0.4, 1.0, 2.0nm |
| Spectral bandwidth | ±0.10nm |
| Wavelength accuracy | ≤0.05nm Wide-spectral-range photomultiplier tube |
| Wavelength | Resolves manganese lamp doublets at 279.5nm and 279.8nm, with trough |
| repeatability | energy between the two lines less than 20%. |
| Detector | 185nm—910nm Auto-adjustment |
| Resolution | 1800 lines/mm Aberration-corrected Czerny-Turner monochromator |
Lighting System
| Lamp holder | Automatic switching for 8 lamp holders |
| Lamp power supply method | 400Hz square wave pulse |
| Lamp current adjustment range | 2-10mA average current |
Atomisation System
| Flame analysis method | |
| Detection limit (Cu) | <0.003μg/ ml |
| Characteristic concentration (Cu) | <0.015μg/ ml/ 1% |
| Burner | 100mm full titanium metal burner |
| Precision | RSD<0.5% |
| Nebuliser | Metal-jacketed high-efficiency glass nebuliser |
| Nebulisation chamber | Corrosion-resistant all-plastic atomisation chamber |
| Safety measures | Features multiple automatic safety protections: acetylene leak alarm, automatic shut-off system, etc. |
| Flame-free (graphite furnace) analytical method | |
| Detection limit (Cd) | <0.4*10ˉ12 g |
| Characteristic quantity (Cd) | <0.5*10ˉ12 g |
| Hydrogenation-atomic absorption method | |
Flame Analysis Method
| Measurement Method | Air-acetylene flame method |
| Concentration Calculation Method | Standard curve method, standard addition method, interpolation method. |
| Number of Repeat | 1st to 99th order, automatic calculation of mean, content, standard |
| Measurements | deviation, relative standard deviation. |
| Result Printing | Print test data or final analysis report. Results exportable to Word |
| worksheets. | |
| Software Environment | Windows XP/Windows 7 operating systems, Chinese professional software. |
Flame-Free (Graphite Furnace) Analytical Method
| Measurement Method | Graphite furnace peak height method, graphite furnace peak area method |
| Concentration Calculation Method | Standard curve method, standard addition method, interpolation method. |
| Background Correction | Deuterium lamp background correction exceeding 60-fold, self-absorption background subtraction exceeding 60-fold |
| Number of Repeat | 1-30 replicates, calculate mean value, provide standard deviation and |
| Measurements | relative standard deviation |
| Result Printing | Print test data or final analytical report. Results exportable to Word worksheets. |
| Software Environment | Windows XP/Windows 7 operating systems, Chinese professional software |
Product Specifications
| Power Supply | AC 220V/50HZ |
| Power | 150VA |
| Net Weight of | 55kg |
| Instrument | |
| Gross Weight of | 80kg |
| Instrument | |
| Instrument Dimensions | 1110mm (width) × 400mm (depth) × 500mm (height) |
| Packaging Dimensions | 1350mm (width) × 600mm (depth) × 850mm (height) |
| Operating Ambient Temperature | 10~35℃ |
| Operating Ambient Humidity | ≤80% |